Description
Top features include:
Triple Junction
- The triple reference design reduces risk of clogging with the reference cell physically separated from the intermediate electrolyte.
Refillable
- The MA918B/1 is refillable with MA9011 3.5M KCl+AgCl electrolyte. This solution is sliver free. Sliver can cause silver precipitate to form at the junction resulting in clogging. Clogging causes erratic and slow readings. The ability to refill the electrolyte also extends the life of the electrode.
Glass Body
- The glass body is easily cleaned and resistant to chemicals.
Conic Tip
- The conic design of the tip provides the largest possible surface area for faster readings and is well suited for testing in many applications.
Technical Specifications
| Body | Glass |
| Reference | Single, Ag/AgCl |
| Junction | Ceramic, triple |
| Electrolyte | 3.5M KCl+AgCl |
| Range | 0-12 pH |
| Tip Shape | Conic |
| Diameter | 0.5 inches |
| Dimensions | 4.7 inches / 6.4 inches |
| Operating Temperature | 23 to 86°F |
| Cable length | 3.3 feet |
| Connection | BNC |
Downloads




BR 18390
US 6244
VN 5061
AR 3389
MX 2273
IQ 1936
IN 1601
ZA 1579
